Bnc 1201 - 6 1/2 Digit Digital Multimeter Uživatelský manuál Strana 79

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 177
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 78
80
Figure 4-14
Remote Interface Operation
Use the following command to make a RATIO measurement.
CONFigure:VOLTage:DC:RATio {<range>|MIN|MAX|DEF},{<resolution>|MIN|MAX|DEF}
4.3.2 % (Percent)
Definition
This mathematical function calculates the ratio of a measurement
reading to a specified target value as the form of percentage. The
calculation formula is shown below:
The specified target value is store in a volatile memory and will be
cleared after the meter has been turned off or a remote interface reset.
How to use % (Percent) function
There are two ways to make a percent measurement: Through the front
panel operation or through the remote interface operation.
Front Panel Operation
To specify a target value, user needs to configure Percent function by
pressing CONFIG + %, then use and to move through the digits