Bnc 1201 - 6 1/2 Digit Digital Multimeter Uživatelský manuál Strana 73

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 177
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 72
74
number of samples per trigger, the number of triggers per event, reading
hold, and the trigger delay for their measurements.
A. Number of samples on each trigger
By default, the multimeter takes only one reading on each trigger, but
you can instruct the multimeter to take specific number (up to 50000) of
readings each time it receives a trigger. The user’s input setting is stored
in a volatile memory which will be cleared after the meter has been
turned off and the default value will be restored. You can set the number
of samples on each trigger through front panel or the remote interface.
Front Panel Operation
Press MENU button and locate “TRIG” submenu by pressing and
buttons. Press ENTER to choose “TRIG” option. Again locate “N SAMPLE”
submenu using and buttons and then press ENTER to select it.
Then set the desired number by pressing and buttons to increase
or decrease the number, and using and buttons to move through
the digits. The locations of these buttons are shown with red rectangle
frames in Figure 4-11.
Procedures: MENU TRIG N SAMPLE <Number>
Figure 4-11
Remote Interface Operation
Use the following command to set the number of samples from your PC
terminal.
+
Zobrazit stránku 72
1 2 ... 68 69 70 71 72 73 74 75 76 77 78 ... 176 177

Komentáře k této Příručce

Žádné komentáře